			 Meant to reply last week but got busy and forgot. 
		
	
		
		
		
		
			
		
		
		
		
	
		
		
	
	
	Mindy, have you performed back to back tests to see if your procedure gives you a different reading than the instructions' procedure would? One test your way and one test the Hanna way? I ask because you're supposed to; Swirl or whatever for 2 minutes, then insert the cuvette and hold the button to start the 3 minute countdown. I double-checked; their online video shows it this way. What I gather is the full countdown allows any un-dissolved powder to settle before it takes the reading. I use the regular model HI713 and find it's always been reliable by doing the above, even with back to back tests - ya know, for those times when it reads 'what the hell? It can't be that high!'  |