Do you have a DI unit hooked up as well? You should be getting around 98% reduction in TDS from the RO membrane alone. Given our tap water is around 22 ppm (that's what mine is right now) you should be getting a reading less than 2 or 3 if your RO unit is set up properly. And then the DI should definitely be getting things down to 0.

I had a problem with a home drinking unit that had a "post carbon" canister plumbed in after the RO membrane to add some TDS back in for palatability (taste) for human consumption. I was getting higher TDS readings than I expected. I ended up changing the order of the canisters to have two carbon filters plumbed consecutively before the RO membrane.

What unit are you using exactly? Unless you have a situation like the one I described above (with a "post carbon" canister) then something is amiss with your RO unit.
__________________
SPS Dedicated 24x24x20 Trimless Tank | 20 g Sump | Bubbble King Mini 160 Protein Skimmer w/ Avast Swabbie | NP Biopellets in TLF Phosban Reactor | ATI Sunpower 6 x 24W T5HO Fixture | EcoTech Vortech MP20 | Modified Tunze Nanostream 6025 | Eheim 1260 Return Pump | GHL Profilux Standalone Doser dosing B-Ionic | Steel Frame Epoxy Coated Stand with Maple Panels embedded with Neodymium Magnets